The G-Series GM50A013101RMM020 Mass Flow Controller features a full scale flow range of 10 sccm and is calibrated for Nitrogen (N2) gas. It has a Swagelok 4 VCR male fitting and 15-pin type D connector with tied grounds for 0 to 5 VDC mass flow control. The normally closed valve features a metal seal. G Series mass flow controllers utilize the latest MKS control algorithms for fast and repeatable response to set point throughout the device control range. Typical response times are on the order of 500 milliseconds. Digital calibration yields 1% of set point accuracy on the nitrogen calibration gas.
±1% of set point for 20 to 100% Full Scale, ± 0.2% of Full Scale for 2 to 20% Full Scale
Repeatability
± 0.3% of Reading
Resolution
0.1% of Full Scale
Zero Temperature Coefficient
<0.05% of Full Scale/°C
Span Temperature Coefficient
<0.08% of Reading/°C
Inlet Pressure Coefficient
<0.02% of Reading/psi
Typical Settling Time
<750 msec, typical above 5% Full Scale
Warm-up Time
30 minutes
Operating Temperature Range
10 to 50°C
Storage Humidity
0 to 95% Relative Humidity, non-condensing
Storage Temperature Range
-20 to 80°C
External Leak Integrity
<1 x 10-09 (scc/sec He)
Leak Integrity Through Closed Valve
Up to 10K valve <0.1% of FS at 40 psig to atmosphere, 20K - 50K valve <1.0% of FS at 40 psig to atmosphere
Wetted Materials
316L S.S. VAR, 316 S.S., Elgiloy®, Nickel, KM45
Seal Material
Metal
Surface Finish
10µ inch average Ra (electropolished)
Weight
3 lbs (1.4 kg)
CE Certified
Yes
Series
G Series
Type
Mass Flow Controller
Calibration Gas
Nitrogen (N2)
Full Scale Flow Range
10 sccm
Maximum Inlet Pressure
150 psig
Normal Operating Pressure Differential
10 to 40 psid
Proof Pressure
1000 psig
Burst Pressure
1500 psig
Fitting Type
4 VCR Male
Control Range
2% to 100% of Full Scale
Flow Input Output Signal
0 to 5 VDC
Connector Type
15-pin Type D Male
Typical Accuracy
±1% of set point for 20 to 100% Full Scale, ± 0.2% of Full Scale for 2 to 20% Full Scale
Repeatability
± 0.3% of Reading
Resolution
0.1% of Full Scale
Zero Temperature Coefficient
<0.05% of Full Scale/°C
Span Temperature Coefficient
<0.08% of Reading/°C
Inlet Pressure Coefficient
<0.02% of Reading/psi
Typical Settling Time
<750 msec, typical above 5% Full Scale
Warm-up Time
30 minutes
Operating Temperature Range
10 to 50°C
Storage Humidity
0 to 95% Relative Humidity, non-condensing
Storage Temperature Range
-20 to 80°C
External Leak Integrity
<1 x 10-09 (scc/sec He)
Leak Integrity Through Closed Valve
Up to 10K valve <0.1% of FS at 40 psig to atmosphere, 20K - 50K valve <1.0% of FS at 40 psig to atmosphere
Wetted Materials
316L S.S. VAR, 316 S.S., Elgiloy®, Nickel, KM45
Seal Material
Metal
Surface Finish
10µ inch average Ra (electropolished)
Weight
3 lbs (1.4 kg)
CE Certified
Yes
GM50A013101RMM020 - Drawings & CAD
Additional CAD file downloads are not available for this product.
GM50A013101RMM020 - Drawings
Additional drawings are not available for this product.
GM50A013101RMM020
Sign In
Request Resource
In order to provide better service and products, please provide the following brief information.
Password Reset
Enter your email address below to reset your account password.
Password Reset
Email Verification Required
Cart Items Updated
Remove Product
Remove this product from your comparison list?
Check Order Status
Provide an order number and postal code to check the status of an order or download an invoice for an order that has shipped. Login to view your complete order history.
Sign In Required
To access this and other valuable technical resources, please sign in or register for a new online account.
Add to Cart
Mass Flow Controller, 10 sccm, 4 VCR Male, 0-5 VDC, DB15, Metal Seal